200S 59
II - 5 Remote Switch Data Output
Example 2: Rotary axis with degrees decimal display
C = + 1260.0000°
1 Coordinate axis
2 Equal sign
3 +/– sign
4 4 to 8 places before the decimal point
5 Decimal point
6 0 to 4 places after the decimal point
7 Blank space
8W for angle (in distance-to-go display: w
9 Carriage return
10 Blank line (Line Feed)
Example 3: Rotary axis with degrees/minutes/seconds display
C = + 360° 23' 45''
1 Coordinate axis
2 Equal sign
3 +/– sign
4 3 to 8 places degrees
5 Colon
6 0 to 2 places minutes
C
=+ 1 2 6 0 . 0 0 0 0 W <CR> <LF>
123456789 10
C
=+3 6 0 : 2 3 : 4 5 W <CR> <LF>
123 4 5 6 7 8 910 1 2